![]()
|
|
| ![]()
|
|
All Dunham & Morrow magnetic locators manufactured after July 1, 2010 have no known inherent problems. However, that does not preclude “Operator Error” problems that fall into the categories of corrosion or broken electronics case. Corrosion Corrosion specifically refers to a process involving the deterioration or degradation of metal components. Corrosion processes are usually electrochemical in nature; when metal atoms are exposed to water molecules in the presence of an electrical charge, they can give up electrons and become positively charged ions. Whenever the customer submerges the instrument in water, all the requirements for metallic corrosion are satisfied and the following pictures show the resulting damage.
Repairing the damage caused by corrosion can cost anywhere from $215 to $375 depending upon the extent of the corrosion damage. In minor cases, we replace the control panel, the battery box terminals, the speaker, and all interconnecting cables. If the corrosion extends to the main PC Board, the cost runs considerably higher. Broken electronics case In some instances, this problem occurs by accident. The instrument was placed on the ground while the surveyor or technician was performing another task, and someone ran over it with the truck. In other cases, we have found that the operator broke the case when he applied excessive torque; usually while in the process of exposing the target he had just detected. The electronics case consists of two custom machined parts that can only be procured as a set. The main cause of this problem is again “Operator Error” and the customer should be billed accordingly. Other problems that can occur In order to properly diagnose performance problems within the unit, you will need to access the instrument main PC Board. You can do that by removing the four adhesive backed rubber bumpers, and then remove the screws underneath and lift-off the cover. Unit is dead or there is no audio output Check that the batteries are installed properly and that the combined battery voltage is above 5.0 VDC. Check that the Power Switch is in the ON position and that all cables are intact and properly connected.
The unit goes into saturation at turn-on The basic locator circuit design employs multiple high forward gain, negative-feedback circuit blocks. In order for the DML2000 magnetic locator to perform correctly, the excitation circuitry must be working properly along with all 4 magnetometer signal channels. Even marginal performance by any one of these circuit blocks can cause the unit to go into saturation when it is turned on. To alleviate this type of problem, we have significantly upgraded the circuit design. Examine the white label on the instrument PC Board, if the designation reads “REV S” or higher, your unit has the new, improved circuit design. The magnetometer excitation frequency has been set to a fixed frequency. In addition, seven other PCB components have been changed; including the four signal channel tuning capacitors, the two excitation inductors and the main frequency setting resistor. The beauty of this fix is that it works, the new 102028 REV S circuit board, works on old or new units, without exception. It is our recommendation that whenever you encounter a unit that goes into saturation at turn-on, you replace the PCB with a new “REV S” PCB.
Figure 1, DML2000 PC Board layout The instrument screams at all times There are two separate potential causes for this type of problem:
To confirm that you have a broken wire in the sensor assembly, connect a DC voltmeter between ground and the PC board signal locations listed below.
If the voltage on the DC voltmeter is nominally ± 4.0 V dc, then that signal channel is in saturation. In a working signal channel, the output will not exceed 3.5 V dc and the output voltage will vary linearly as the instrument sensor assembly is moved around in the Earth’s magnetic field. If the broken wire is in the excitation circuit, then two of the signal channels will be in saturation. To fix a broken wire in the sensor assembly follow these procedures:
In case you did not notice, the sensor connector goes down as you face the unit (it is on the underside of the sensor PC board). Push the sensor assembly back into position, and then hold it in the tube as you reattach the Electronics PC board to the other end. Seal the sensors with Silastic or Room Temperature Vulcanizer (RTV) and then glue a replacement crutch tip onto the end of the sensor tube. Click here For a list of available replacement parts. Other problems This covers all of the typical problems. If this does not cover your problem call our service department at 703-661-2144 or email warren@magneticlocator.com. Always remember that the
ultimate responsibility for any repair is Dunham & Morrow, Inc. and
you can send the unit back to us for repair at anytime. |
||||
|
Dunham & Morrow, Inc., 43676 Trade Center Place, Suite 145,
Dulles, VA 20166
Site
design by
Miller Creative! |
|